Tel:+86 512 62525801
The quality control involves gravimetric testing of each pipette with distilled water at 22℃.
Features for MicroPette
- Lightweight, ergonomic, low force design
- Digital display clearly reads volume setting
- The pipettes cover volume range of 0.1μl to 10ml
- Easy to calibrate and maintain with tool supplied
- Design helps avoid repetitive strain injuries
- Calibrated in accordance with ISO8655. Each pipette supplied with individual test certificate
- The low part is available for autoclaving
- Manufactured from innovative materials
Features for multi-channel MicroPette
- 8 and 12 channel pipettes are available for standard 96-well plate
- Dispensing head rotates for optimum pipetting convenience
- Individual piston and tip cone assemblies allow easy repair and maintenance
- Compound material tip cone design allows visual seal verification
- Can be used with universal style pipette tips
Specifications
Single-channel Adjustable Volume Pipettes |
|||||||||||
Volume Range |
Increment |
Test Volume |
Maximum permissible systematic error (Inaccuracy) |
Maximum permissible random error (Imprecision) |
|||||||
% |
μL |
% |
μL |
||||||||
0.1-2.5μL |
0.05μL |
2.5μL |
2.50% |
0.0625 |
2.00% |
0.05 |
|||||
1.25μL |
3.00% |
0.0375 |
3.00% |
0.0375 |
|||||||
0.25μL |
12.00% |
0.03 |
6.00% |
0.015 |
|||||||
0.5-10μL |
0.1μL |
10μL |
1.00% |
0.1 |
0.80% |
0.08 |
|||||
5μL |
1.50% |
0.075 |
1.50% |
0.075 |
|||||||
1μL |
2.50% |
0.025 |
1.50% |
0.015 |
|||||||
2-20μL |
0.5μL |
20μL |
0.90% |
0.18 |
0.40% |
0.08 |
|||||
10μL |
1.20% |
0.12 |
1.00% |
0.1 |
|||||||
2μL |
3.00% |
0.06 |
2.00% |
0.04 |
|||||||
5-50μL |
0.5μL |
50μL |
0.60% |
0.3 |
0.30% |
0.15 |
|||||
25μL |
0.90% |
0.225 |
0.60% |
0.15 |
|||||||
5μL |
2.00% |
0.1 |
2.00% |
0.1 |
|||||||
10-100μL |
1μL |
100μL |
0.80% |
0.8 |
0.15% |
0.15 |
|||||
50μL |
1.00% |
0.5 |
0.40% |
0.2 |
|||||||
10μL |
3.00% |
0.3 |
1.50% |
0.15 |
|||||||
20-200μL |
1μL |
200μL |
0.60% |
1.2 |
0.15% |
0.3 |
|||||
100μL |
0.80% |
0.8 |
0.30% |
0.3 |
|||||||
20μL |
3.00% |
0.6 |
1.00% |
0.2 |
|||||||
50-200μL |
1μL |
200μL |
0.60% |
1.2 |
0.15% |
0.3 |
|||||
100μL |
0.80% |
0.8 |
0.30% |
0.3 |
|||||||
50μL |
1.00% |
0.5 |
0.40% |
0.2 |
|||||||
100-1000μL |
5μL |
1000μL |
0.60% |
6 |
0.20% |
2 |
|||||
500μL |
0.70% |
3.5 |
0.25% |
1.25 |
|||||||
100μL |
2.00% |
2 |
0.70% |
0.7 |
|||||||
200-1000μL |
5μL |
1000μL |
0.60% |
6 |
0.20% |
2 |
|||||
500μL |
0.70% |
3.5 |
0.25% |
1.25 |
|||||||
200μL |
0.90% |
1.8 |
0.30% |
0.6 |
|||||||
1000-5000μL |
50μL |
5000μL |
0.50% |
25 |
0.15% |
7.5 |
|||||
2500μL |
0.60% |
15 |
0.30% |
7.5 |
|||||||
1000μL |
0.70% |
7 |
0.30% |
3 |
|||||||
2-10mL |
0.1mL |
10mL |
0.60% |
60 |
0.20% |
20 |
|||||
5mL |
1.20% |
60 |
0.30% |
15 |
|||||||
2mL |
3.00% |
60 |
0.60% |
12 |
|||||||
8-channel Adjustable Volume Pipettes |
|||||||||||
Volume Range |
Increment |
Test Volume |
Maximum permissible systematic error (Inaccuracy) |
Maximum permissible random error (Imprecision) |
|||||||
% |
μL |
% |
μL |
||||||||
0.5-10μL |
0.1μL |
10μL |
1.50% |
0.15 |
1.50% |
0.15 |
|||||
5μL |
2.50% |
0.125 |
2.50% |
0.125 |
|||||||
1μL |
4.00% |
0.04 |
4.00% |
0.04 |
|||||||
5-50μL |
0.5μL |
50μL |
1.00% |
0.5 |
0.50% |
0.25 |
|||||
25μL |
1.50% |
0.375 |
1.00% |
0.25 |
|||||||
5μL |
3.00% |
0.15 |
2.00% |
0.1 |
|||||||
50-300μL |
5μL |
300μL |
0.70% |
2.1 |
0.25% |
0.75 |
|||||
150μL |
1.00% |
1.5 |
0.50% |
0.75 |
|||||||
50μL |
1.50% |
0.75 |
0.80% |
0.4 |
|||||||
12-channel Adjustable Volume Pipettes |
|||||||||||
Volume Range |
Increment |
Test Volume |
Maximum permissible systematic error (Inaccuracy) |
Maximum permissible random error (Imprecision) |
|||||||
% |
μL |
% |
μL |
||||||||
0.5-10μL |
0.1μL |
10μL |
1.50% |
0.15 |
1.50% |
0.15 |
|||||
5μL |
2.50% |
0.125 |
2.50% |
0.125 |
|||||||
1μL |
4.00% |
0.04 |
4.00% |
0.04 |
|||||||
5-50μL |
0.5μL |
50μL |
1.00% |
0.5 |
0.50% |
0.25 |
|||||
25μL |
1.50% |
0.375 |
1.00% |
0.25 |
|||||||
5μL |
3.00% |
0.15 |
2.00% |
0.1 |
|||||||
50-300μL |
5μL |
300μL |
0.70% |
2.1 |
0.25% |
0.75 |
|||||
150μL |
1.00% |
1.5 |
0.50% |
0.75 |
|||||||
50μL |
1.50% |
0.75 |
0.80% |
0.4 |
|||||||
Fixed Volume Pipettes |
|||||||||||
Volume Range |
Increment |
Test Volume |
Maximum permissible systematic error (Inaccuracy) |
Maximum permissible random error (Imprecision) |
|||||||
% |
μL |
% |
μL |
||||||||
5μL |
- |
5μL |
1.3% |
0.065 |
1.2% |
0.06 |
|||||
10μL |
- |
10μL |
0.8% |
0.08 |
0.8% |
0.08 |
|||||
20μL |
- |
20μL |
0.6% |
0.12 |
0.5% |
0.1 |
|||||
25μL |
- |
25μL |
0.5% |
0.125 |
0.3% |
0.075 |
|||||
50μL |
- |
50μL |
0.5% |
0.25 |
0.3% |
0.15 |
|||||
100μL |
- |
100μL |
0.5% |
0.5 |
0.3% |
0.3 |
|||||
200μL |
- |
200μL |
0.4% |
0.8 |
0.2% |
0.4 |
|||||
250μL |
-- |
250μL |
0.4% |
1.0 |
0.2% |
0.5 |
|||||
500μL |
- |
500μL |
0.3% |
1.5 |
0.2% |
1.0 |
|||||
1000μL |
- |
1000μL |
0.3% |
3.0 |
0.2% |
2.0 |
|||||
2000μL |
- |
2000μL |
0.3% |
6.0 |
0.15% |
3.0 |
|||||
5000μL |
- |
5000μL |
0.3% |
15 |
0.15% |
7.5 |